Description

UreaseTest Package Insert

Unit: 1 Box

Regulatory Status: CLIA Waived

Contents: 3mL dropper bottle

Storage Requirements: 59° – 86° F

QUALITY CONTROL: We recommend a positive external control must be preformed when opening a new test

kit. The test kit size must not exceed 50. Accutest H. pylori Urease Test controls come in two forms: Dry Paper

Dot Control and Liquid Control. Either may be used.

If the Accutest H. pylori Urease Test is negative after 1 hour and there is a question if the test has functioned

properly, perform the following positive control test:

LIQUID CONTROL

1. Peel back the label on the Accutest H. pylori Urease Test so that the yellow ring is visible.

2. Place 1 small drop of control solution directly on the testing surface. The control solution gives a positive result.

3. Reseal the Accutest H. pylori Urease Test.

4. Observe the ring for a positive color change to magenta. The color change should occur within 30 seconds but

may take up to 1 hour. Please Note: In some cases you may notice the center of the test changing to magenta

before the outer ring. This is NORMAL and caused by the introduction of liquid onto the filter paper. Any change in

color from yellow to magenta on any part of the test surface should be considered POSITIVE.

5. If after 1 hour there is no change in the color, please contact Jant Pharmacal Corporation Customer Service at

800-676-5565.

DRY PAPER DOT CONTROL

1. Peel back the label on the Accutest H. pylori Urease Test so that the yellow ring is visible.

2. Place 1 – 2 ‘Accutest H. pylori Urease Test ‘Positive Control Dots on the ring with one drop of distilled water.

The Control Dot gives a positive result.

3. Reseal the Accutest H. pylori Urease Test.

4. Observe the ring for a positive color change to magenta. The color change should occur in about 5 minutes but

may take up to 1 hour. Please Note: In some cases you may notice the center of the test changing to magenta

before the outer ring. This is NORMAL and caused by the introduction of liquid onto the filter paper. Any change in

color from yellow to magenta on any part of the test surface should be considered POSITIVE.

5. If after 1 hour there is no change in the color, the test is Invalid. Please contact Jant Pharmacal Corporation

Customer Service at 800-676-5565.