Description
![]()
Contents: Afinion™ ACR Control C I: Citrate buffer with albumin and creatinine, (1) 1.0mL Afinion™ ACR Control C II: Citrate buffer with albumin and creatinine. Package Insert
For Use With: Afinion 2 Analyzer
Storage Requirements: 36° – 46° F
Quality control testing using the Afinion HbA1c Control should be done to confirm that your Afinion Analyzer System is working properly and provides reliable results. Only when controls are used routinely and the values obtained are within acceptable ranges can accurate results for patient samples be assured.
Frequency:
- With each new shipment of Afinion ACR Test Kits.
- With each new lot of Afinion ACR Test Kits.
- At least every 30 days.
- When training new operators in correct use of Afinion ACR Test Kit and Afinion Analyzer.
- Anytime an unexpected test result is obtained (section “Interpretation of results”, Afinion ACR Package Insert).
- If local, state and/or federal regulations require more frequent testing of control materials, quality control should be performed in compliance with these regulations.