Accutest H. Pylori Urease Control

$40.00

Special Order

SKU: IDC73 Category:

Description

UreaseTest Package Insert

Unit: 1 Box
Regulatory Status: CLIA Waived
Contents: 3mL dropper bottle
Storage Requirements: 59° – 86° F

QUALITY CONTROL: We recommend a positive external control must be preformed when opening a new test
kit. The test kit size must not exceed 50. Accutest H. pylori Urease Test controls come in two forms: Dry Paper
Dot Control and Liquid Control. Either may be used.
If the Accutest H. pylori Urease Test is negative after 1 hour and there is a question if the test has functioned
properly, perform the following positive control test:

LIQUID CONTROL
1. Peel back the label on the Accutest H. pylori Urease Test so that the yellow ring is visible.
2. Place 1 small drop of control solution directly on the testing surface. The control solution gives a positive result.
3. Reseal the Accutest H. pylori Urease Test.
4. Observe the ring for a positive color change to magenta. The color change should occur within 30 seconds but
may take up to 1 hour. Please Note: In some cases you may notice the center of the test changing to magenta
before the outer ring. This is NORMAL and caused by the introduction of liquid onto the filter paper. Any change in
color from yellow to magenta on any part of the test surface should be considered POSITIVE.
5. If after 1 hour there is no change in the color, please contact Jant Pharmacal Corporation Customer Service at
800-676-5565.
DRY PAPER DOT CONTROL
1. Peel back the label on the Accutest H. pylori Urease Test so that the yellow ring is visible.
2. Place 1 – 2 ‘Accutest H. pylori Urease Test ‘Positive Control Dots on the ring with one drop of distilled water.
The Control Dot gives a positive result.
3. Reseal the Accutest H. pylori Urease Test.
4. Observe the ring for a positive color change to magenta. The color change should occur in about 5 minutes but
may take up to 1 hour. Please Note: In some cases you may notice the center of the test changing to magenta
before the outer ring. This is NORMAL and caused by the introduction of liquid onto the filter paper. Any change in
color from yellow to magenta on any part of the test surface should be considered POSITIVE.
5. If after 1 hour there is no change in the color, the test is Invalid. Please contact Jant Pharmacal Corporation
Customer Service at 800-676-5565.

Additional information

Weight 1 lbs
Dimensions 4 × 4 × 4 in

You may also like…